Frame Modifications

Heel Aperture

Central frame opening beneath a continuous top cover, left without a plug. Creates local structural relief; check rim pressure, clearance and the surrounding heel’s tolerance.

Viscoelastic insert within a central heel-frame aperture. Provides a compliant covered heel seat while surrounding frame contact remains; monitor perimeter pressure and plug compression.

Heel Aperture, view 1
  • mm

An unfilled heel aperture removes a central piece of the frame beneath the heel. The top cover spans the opening, while the surrounding heel seat retains support. It changes the central interface from direct structural contact to a covered relief region.

A circular opening, approximately 25–30 mm in diameter in the published construction, is made in the central heel frame. In this version it remains without a viscoelastic plug. The continuous top cover covers the opening, so the foot does not meet a bare frame rim. The surrounding cup and the underside assembly remain part of the design.

The covered center has less direct frame support and can deflect into the available space. Surrounding heel tissue receives more of the supporting contact. If the cover reaches a hard post or shoe surface immediately below it, or if the rim concentrates pressure, the expected relief can be lost. Related aperture studies support the concept of focal space, rather than proving this exact heel construction [E06, E10].

Consider it for a focal central heel-contact problem when the surrounding tissue can accept the redistributed load and the covered opening remains useful. The therapeutic goal is reduced troublesome direct contact. A radiographic heel spur alone does not identify the pressure-sensitive footprint or prove the need for an aperture.

A sensitive or thinned heel fat pad may tolerate a broad compliant seat better than an unfilled opening. Rim pressure and bottoming out are the main fitting concerns. Acute calcaneal pain, suspected stress injury or a neuropathic ulcer requires its own assessment and management [E59].

Diameter is separate from frame thickness and top-cover thickness. The approximate 25–30 mm circle is a construction reference, not a universal lesion size. Location and rim tolerance follow the actual heel footprint. The 1.5–6 mm cushion increments do not define an aperture diameter or a depth menu.

A broad heel cushion can soften receiving contact around the opening. Heel-cup containment can help the tissue remain seated. Coordinate the post and filler so a hard base does not eliminate useful clearance beneath the covered center.

A viscoelastic plug provides a compliant central seat instead of an unfilled relief. A heel cushion changes the soft layer while preserving the frame. A heel-spur pad or local sweet spot can serve a differently located contact target.

Assess the central heel, rim and surrounding seat during walking. Check for bottoming out and stable device seating. Follow pain, skin and cover wear; reduced sensation calls for pressure assessment and direct inspection rather than comfort alone.

Central heel apertures apply local-relief logic within a containing frame. Related opening studies explain focal pressure redistribution, and heel-cup work informs the surrounding tissue support [E06, E70]. Clinical evidence for a compliant heel plug should not be presented as proof of an identical unfilled aperture response.

A plugged heel aperture replaces central frame contact with a compliant heel seat. The frame and cup continue to support the surrounding heel while the insert changes the interface at the center. It is a way to combine structure around the heel with cushioning beneath its selected plantar footprint.

A circular central opening, approximately 25–30 mm in diameter in the published construction, receives a viscoelastic plug. The plug occupies the aperture beneath the continuous top cover and blends with the surrounding heel interface. The foot meets a smooth covered surface rather than an exposed hole or hard cut edge.

The plug deforms more readily than the surrounding frame and can redistribute central heel contact. Its material, thickness, footprint and the base beneath it determine how much deformation remains useful. Pressure can move toward the perimeter or elsewhere on the foot, so assess the complete interface. Plug studies measured heel-pressure changes without consistently finding superior comfort or function [E49, E50].

Consider it for central heel-interface sensitivity when a compliant seat is desirable within an otherwise supportive frame. It can participate in plantar-heel-pain or fat-pad-tolerance care alongside the appropriate assessment and rehabilitation. The therapeutic goals are less troublesome contact and better activity tolerance.

A compressed plug can bottom out, while an abrupt material boundary can create rim pressure. The design does not remove an osteophyte or prove recovery of fascial pathology. A high-risk heel needs skin and pressure surveillance, and an active ulcer needs appropriate wound offloading [E59].

Aperture diameter, plug thickness and the overlying cushion are separate dimensions. The plug is matched to the opening and intended loaded seat; more softness or depth is not automatically better. Keep the surface transition smooth and reassess useful compression over time.

A broad cushioning layer can soften the plug perimeter. A suitably shaped heel cup provides surrounding containment. Posts and filler stabilize the base. A fascia groove may address a separate arch-band irritation rather than the central heel contact.

An unfilled aperture provides space without a central insert. A heel cushion preserves the continuous frame and adds a softer interface above it. A broader accommodative device may be preferable when the entire heel pad is vulnerable.

Assess central heel pressure, perimeter contact and device stability while walking. Follow pain and function as well as the mechanical result. Inspect compression and material condition; replace the plug when it no longer provides the intended compliant seat.

The plug makes the local-relief approach more continuous at the foot-facing surface. Direct studies of heel-plug assemblies support pressure redistribution, but a clinical comparison did not show corresponding superiority in every patient outcome [E49, E50]. That is a useful distinction between a measured interface effect and a treatment guarantee.

  1. The effect of removing plugs and adding arch support to foam based insoles on plantar pressures in people with diabetic peripheral neuropathy.

    Twenty-six adults with diabetic neuropathy. Removing foam-insole plugs at the highest-pressure forefoot region reduced mean peak pressure by 32.3%; adding arch support produced further offloading.

    Removing local material reduced pressure in a neuropathic population. The construct differs from a simple circular punch and does not alone establish active-ulcer healing.

  2. Preliminary investigation on the reduction of plantar loading pressure with different insole materials (SRP--Slow Recovery Poron, P--Poron, PPF--Poron +Plastazote, firm and PPS--Poron+Plastazote, soft).

    Five subjects tried several 6.4 mm material combinations and felt with a first-MTP aperture. The aperture reduced first-MTP peak pressure by 37%; firm Poron–Plastazote reduced it by 29%.

    A very small pilot supports local material and aperture principles. Its numerical results belong to the tested felt and foam constructs.

  3. Comparison of custom and prefabricated orthoses in the initial treatment of proximal plantar fasciitis.

    Two hundred thirty-six patients used stretching alone or stretching plus one of four inserts. At eight weeks, improvement was reported by 95% with silicone, 88% with rubber, 81% with felt and 68% with custom orthoses.

    Every group used stretching. The tested inserts differ from the covered Myolite cushion and horseshoe shapes in this guide.

  4. Custom-made foot orthoses with and without heel plugs and their effect on plantar pressures during treadmill walking.

    Ten healthy adults tried otherwise-matched custom orthoses with or without a soft Poron heel plug. The plug reduced hindfoot pressure while increasing some midfoot pressures.

    The plug is a related focal heel accommodation. Pressure transfer in healthy adults does not establish clinical symptom benefit.

  5. Custom-made foot orthoses with and without heel plugs and their effect on treatment outcomes and plantar pressures in patients with plantar fasciitis: A crossover study.

    Twenty-one patients with plantar fasciitis tried orthoses with and without heel plugs. Plugs reduced hindfoot pressure, but function and comfort did not differ between designs after the treatment periods.

    Lower hindfoot pressure did not produce better function or comfort between designs. The paper was published online in 2025 and in a 2026 journal issue.

  6. Guidelines on offloading foot ulcers in persons with diabetes (IWGDF 2023 update).

    The IWGDF 2023 update recommends a non-removable knee-high offloading device first for suitable neuropathic plantar forefoot or midfoot ulcers, with alternatives when contraindicated or not tolerated.

    The guideline concerns active diabetes-related ulcers. It defines when wound offloading takes priority over routine in-shoe padding; the journal issue is dated 2024.

  7. The relationship of the position of the metatarsal heads and peak plantar pressure.

    CT and pressure measurements showed that first-, fourth- and fifth-metatarsal-head positions did not coincide precisely with local peak pressure.

    Anatomical location and the pressure-sensitive footprint are related but distinct. Accurate fitting can require both rather than a bony landmark alone.

  8. Treating Heel Pain in Adults: A Randomized Controlled Trial of Hard vs Modified Soft Custom Orthotics and Heel Pads.

    Forty-four adults with heel pain received hard or modified soft custom devices. Pain intensity and interference improved over time, with no difference between the groups and no overall improvement in the function measure. Soft devices cost less and needed fewer fabrication visits.

    Greater rigidity is not automatically greater symptom relief. Both treatment groups received an intervention, so improvement over time does not establish superiority to no treatment or isolate a reinforcement feature.

  9. Changes in calcaneal pitch and heel fat pad thickness in static weight bearing radiographs while wearing shoes with arch support and heel cup orthotics.

    Twenty-four feet in 12 female soccer players were examined with and without an arch-support and heel-cup insole. Mean calcaneal pitch increased 1.05° and heel fat-pad thickness increased 1.25 mm. Nine participants rated the overall experience good or excellent.

    Heel containment can alter soft-tissue geometry under static load. A small study of a complete insole does not prove that a particular cup depth reduces pain or provides the same effect during walking.

  • HA-NO Heel Aperture No Plug
  • HA-PL Heel Aperture With Visco Plug